PRS-812
Resistance Meter Set Specifications
Portable, battery operated ohmmeter
Overview | Specifications
SPECIFICATIONS:
Range:
Optimal
Resistance Range from <0.1 (1.0E-1) Ohms to 1 Tera
ohms (1.0E+12 ohms). Maximum resistance indication: 200 Tera
ohms (2.0E+14 ohms).
Test Voltages:
Automatic mode Default: <0.01 to 10 volts Variable 1.0E-1
to 1.0E+4 Ohms
Constant Voltage 10 volts ± <0.2 volts 1.0E+4 to 1.0E+6
Ohms
At 10 & 100 Volts 100 volts ± <2.0 volts 1.0E+6
to 2.0E+14 Ohms
Manual Mode:
< 0.01 to 10 volts Variable 1.0E-1 to 1.0E+4 Ohms
10 volts ± <0.2 volts 1.0E+2 to 1.0E+9 Ohms
100 volts ± <2.0 volts 1.0E+5 to 2.0E+14 Ohms
Accuracy:
Overall: ± <5% at ambient conditions (at 23°F and 30%
Rh).
Specific Range Tolerances:
1.0E-1 to 1.0E+1 ohms: ±5% Corrected for Test Lead Resistance
1.0E+1 to 1.0E+12 ohms: ±2.0% with 10-foot Test Leads
1.0E+12 to 2.0E+14 ohms: >±50%
Display:
Multi-function 2-5/8” x 1-5/8” Liquid Crystal Display
with 0.5" digit height. Displays 3-1/2 digits in Ohms, or
1.0EXX in exponential format. Ohms Display indicators: W, KW,
MW, GW and TW. Includes 19-segment analog scale (1-10 with 0.5
indication) with x1, x10, & x100 multipliers. Number of Data
Points in Memory (0 – 80). Automatic Electrification Time
(seconds), or Time required to Manually obtain steady state measurement.
Displays data HOLD, BATTERY status, MIN, MAX, AVG, REC and Test
Voltage (<10, 10, or 100 V)
LED Indicators:
14 Color LED’s from <103 to >1014 ohms. Colors (RED,
GREEN, YELLOW, or Blank/OFF).
Timer Memory:
Times measurements In Seconds up to 99 seconds (Displayed on
LCD)
Register stores up to 80 data points (MEM # Displayed after RESET)
Response & Electrification:
Response from >0.1 to <1.0E06 Ohms: <2.0 seconds
Average Measurement Period from 0.1 ohms to 1.0E12 Ohms 2.5 Seconds.
Calculated Electrification Period per ESD S11.11, 7.5 seconds
0.1 ohms to 10E+12 Ohms. Programmed Electrification 1.0E+12 to >2.0E14
Ohms: 15.0+ seconds
Power:
Two 9-VDC alkaline batteries. Nominal battery life 25 hours.
Dimensions:
4.0" wide x 6.0" long x 2.0" deep.
Weight:
22 ounces, with batteries
Open Circuit Current (I):
< 4ma @ 100 Volts
Function/ Mode:
Toggles Through Six (6) Operation Modes:
Resistance Range Select:
2 Triangular button Arrows: UP and DOWN ¯. Select
Resistance
Range in single decades in Manual and Automatic/Manual modes
Test Volts:
Manual Selection of <10, 10 or 100 volts in Manual Mode
Record/Recall:
Turns Memory Register ON if OFF. Calculates and Displays Minimum,
Maximum and Average of data stored in Memory Register.
Clear:
Erases data in Memory Register; if in HOLD mode, discards most
recent Held Value
ON/OFF:
Power-up, perform functional & Battery tests. Power down
if ON.
Batt. Test:
Displays GOOD on LCD if acceptable voltage or Lo if unacceptable
Reset:
Enters (saves) data into Memory Register, Clears HOLD and Display.
Test:
Begins measurement sequence.
Battery Buss Cut OFF:
ON/OFF Switch isolates batteries from instrument circuits for
storage & transport
Automatic Function and Exceptional Accuracy:
The PRS-812’s Automatic mode is designed to simplify ESD
Auditors’ measurement and data recording challenges, while
enhancing measurement reproducibility and accuracy. During power-up,
the instrument automatically checks its measurement functions,
multiple displays and performs a Battery Test. Once “GOOD” is
displayed, the operator simply presses the TEST button to obtain
an accurate measurement. The instrument automatically selects
and adjusts test voltage, resistance range, electrification period,
then displays and Holds the measurement. The displayed measurement
is the average of eight consecutive measurements, all within ±5%
of each other. The measurement is “saved” in the
Memory Register by pressing RESET. The instrument is now ready
for the next measurement.
Data Log & Calculation:
In RECORD mode, the PRS-812 stores up to 80 measurements, and
on demand will calculate and display the Minimum, Maximum and
Average measurements stored in the register.
Technical Documents:
» PRS-812 Data Sheet |